mks-u
Multipurpose dosimeter-radiometer
*Produced in two versions: for Multi-Purpose use; for Fire Safety Professionals.
Certificates
  In Ukraine:
Pattern Approval Certificate of Measuring Instruments # UA-MI/1-918-2001
of January 12, 2001 (Ó1355-00 number in the State Register for Measuring Instruments)
Conformance Certificate of Measuring Instruments with the Approved Pattern
# UA-MI/2-1003-2003 of February 27, 2003
Hygienic conclusion of the State Sanitary-Hygienic Expertise # Â-7.03/140
of October 24, 2000
ÒÓ Ó 22362867.005-2000.
ÒÓ Ó 33.2-22362867-005-01-2003 (military accepted).
  Overseas:
Pattern Approval Certificate of Measuring Instruments # 3568 of April 19, 2007 (KZ.02.03.01878-2007/Ó1355-00 number in the State Register for Measuring Instruments), Kazakhstan
Photo
3d-ico
Multipurpose dosimeter - radiometer
Purpose of use
Measurement of gamma and X-ray radiation equivalent dose rate.
Measurement of gamma and X-ray radiation equivalent dose.
Measurement of surface beta-particles flux density.
Measurement of gamma and X-ray radiation equivalent dose accumulation time.
Measurement of surface alpha-particle flux density with the help of the BDPA-07 detecting unit of alpha radiation (on demand).
Measurement of thermal and fast neutrons flux density with the help of the BDPN-07 detecting unit (on demand).
Features
Automatic setting of measurement intervals and ranges.
Audio signalling of each detected gamma-quantum or beta-particle.
Digital display and control keys backlight if operating in the dark.
Storage battery (five nickel-cadmium AA size batteries) charging with the built-in charger from:
- integral heliobattery;
- 12V automobile battery;
- 220V/50Hz industrial network by voltage converter.
Multilevel indication of power supply discharge.
Option to operate under atmospheric precipitation, dusty conditions and at submerge of the remote gamma radiation detector in the water at 0.5 m depth.
Measurement of emergency EDR levels of gamma radiation by placing the remote detector at up to 30 m distance.
Analog indicator of radiation intensity.
Up to 4096 measurement results recording in the nonvolatile memory with further transfer to the computer through infrared port.
Viewing of the recorded measurement results on the personal digital display.
Option to operate in the individual protection (rubber gloves).
Wide operating temperature range (-40...+50°C).
Digital display resistance to the temperature of +95°C.
The device includes:
gas-discharge Geiger-Muller gamma radiation detectors without return run of counting response;
silicon beta radiation detector;
emergency gamma radiation detector of "SELDI" type (scintillator-photodiode).
Specifications
Measurement of gamma radiation parameters:
- Measurement range of equivalent dose rate by the combined detecting unit 0.1µSv/h…10Sv/h
- Measurement range of equivalent dose rate by the remote detector Sv/h 0.01…100
- Measurement range of equivalent dose by the detector built-in the control panel mSv 0.001…9 999
- Basic relative errors of EDR and ED measurement with 137Cs calibration:
- for built-in and combined detecting unit ±15%
- for remote detector: ±15%
- Energy range of measurement MeV 0.05…3.0
- Equivalent dose accumulation time range and accuracy of measurement 1min…100h
±1min per 100h
Measurement of beta radiation parameters:
- Surface beta-particles flux density measurement range 1/(cm²·min) 10…200 000
- Energy range of measurement MeV 0.3…3.0
- Basic relative error of beta-particles flux density measurement with (90Sr+90Y) calibration. ±20%
Measurement time intervals seconds 2…50
Storage battery life hours 100
Operating temperature range
(for digital display)
°C -40…+50
(-40…+95)
Weight of the delivery kit in packing kg 8.0
Dimensions of the delivery kit in packing mm 490×255×130
Complete delivery kit
-
-
-
-
-
-
-
-
-
-
-
-



-
-
control panel with non-detachable combined gamma, beta detecting unit;
heliobattery;
short telescopic bar for the detecting unit with a clip for belt fixation;
12V cable (10m) used for charging from the automobile battery;
220-12V adapter;
nickel-cadmium batteries (5 pcs, AA size);
headphones;
holder with the control radiation source (137Cs);
packing case;
operating manual;
logbook;
emergency detecting unit kit:
- remote gamma radiation detector with 30 m cable;
- telescopic bar (5m) with a cable hanger for cable winding;
- telescopic bar case;
spare parts;
exchange infrared port adapter and software(at customer's request).
The kit is completed at the customer's request.
Multipurpose dosimeter - radiometer